图文详情
产品属性
相关推荐
SPECIFICATIONS | ||||
MODEL |
7130 |
7132 |
7140 |
7142 |
AC WITHSTAND VOLTAGE | ||||
Output Rating |
5KVAC/20mA | |||
Range |
Resolution |
Accuracy | ||
Output Voltage, KVAC |
0-5.00 |
0.01 |
± ( 2% of setting + 5V ) | |
Output Frequency |
50Hz/60Hz ±100ppm, User Selection | |||
Output Waveform |
Sine Wave ,THD.<2% (Resistive Load), Crest Factor = 1.3 - 1.5 | |||
Output Regulation |
± (1% of output + 5V), From no load to full load | |||
SETTINGS | ||||
Max-Limit AC Current, mA |
0 - 20.00 |
0.01 |
± ( 2% of setting + 2counts ) | |
Min-Limit AC Current, mA |
0 - 9.999 |
0.001 |
± ( 2% of setting + 2counts ) | |
Ramp Up Time, second |
0.1 - 999.9 |
0.1 |
± ( 0.1% + 0.05sec ) | |
Ramp Down Time, second |
0 - 999.9 |
0.1 | ||
Dwell Time, second |
0, 0.3 - 999.9 |
0.1 | ||
(0=continuous) | ||||
Arc Detection |
0, 1 - 9 ranges (0=OFF, 9 is the most sensitivity) | |||
DC WITHSTAND VOLTAGE (7140,7142 ONLY) | ||||
Output Rating |
6KVDC/7500µA | |||
Output Voltage, KVDC |
0 - 6.00 |
0.01 |
± ( 2% of setting + 5V ) | |
Output Ripple |
<5% ( 6KV/7500µA at Resistive Load ) | |||
SETTINGS | ||||
Max-Limit DC Current, µA |
0 - 7500 |
1 |
± ( 2% of setting + 2counts ) | |
Min-Limit DC Current, µA |
0 - 999.9 |
0.1 |
± ( 2% of setting + 2counts ) | |
Ramp Up Time, second |
0.1 - 999.9 |
0.1 |
± ( 0.1% + 0.05sec ) | |
Ramp Down Time, second |
0, 1.0 - 999.9 |
0.1 | ||
Dwell Time, second |
0, 0.4 - 999.9 |
0.1 | ||
(0=continuous) | ||||
Arc Detection |
0, 1 - 9 ranges (0=OFF, 9 is the most sensitivity) | |||
Discharge Time |
< 200 msec | |||
INSULATION R*ISTANCE (7132,7142 ONLY) | ||||
Output Rating |
1KVDC/9999MΩ | |||
Output Voltage, DCV |
30 - 1000 |
10 |
± ( 2% of setting + 5V ) | |
SETTINGS | ||||
Max-Limit Resistance, MΩ |
0, 1 - 9999 (0=OFF) |
1 |
± ( 2% of setting+2counts ) at 500-1000VDC, | |
1 -1000 MΩ | ||||
Min-Limit Resistance, MΩ |
1 - 9999 |
1 |
± ( 5% of setting+2counts ) at 500-1000VDC, | |
1000 -9999 MΩ | ||||
± ( 8% of setting+2counts ) at 30-500VDC, | ||||
1 -1000 MΩ | ||||
Ramp Up Time, second |
0.1 - 999.9 |
0.1 |
± ( 0.1% + 0.05sec ) | |
Ramp Down Time, second |
0, 1.0 - 999.9 |
0.1 | ||
Delay Time, second |
0, 0.5 - 999.9 |
0.1 | ||
(0=continuous) | ||||
MEASUREMENT | ||||
AC/DC Voltage, KV |
0 - 6.00 |
0.01 |
± ( 1.5% of reading + 1count ) | |
DC Voltage, V (IR only) |
30 - 1000 |
1 |
± ( 1.5% of reading + 5V ) | |
AC Current, mA |
0 - 3.500 |
0.001 |
± ( 2% of reading + 2counts ) | |
3.00 - 20.00 |
0.01 | |||
DC Current, µA |
0 - 350.0 |
0.1 |
± ( 2% of reading + 2counts ) | |
DC Current, mA |
0.300 - 3.500 |
0.001 | ||
3.00 - 7.50 |
0.01 | |||
Resistance, , MΩ |
1 - 9999 |
0.001 |
± ( 2% of setting+2counts ) at 500-1000VDC, | |
( Auto Range) |
0.01 |
1 -1000 MΩ | ||
|
0.1 |
± ( 5% of setting+2counts ) at 500-1000VDC, | ||
|
1 |
1000 -9999 MΩ | ||
|
|
± ( 8% of setting+2counts ) at 30-500VDC, | ||
|
|
1 -1000 MΩ | ||
GENERAL | ||||
Input Voltage AC |
115/230Vac±15%, 50/60Hz ± 5%, Fuse 3.15A | |||
PLC Remote Control |
Input : Test , Reset , Memory1.2.3 , Interlock | |||
Output : P* , Fail , Processing , Reset-Out | ||||
Interface(Option) |
RS485 Interface | |||
Memory |
10 memories, 3 steps/memory | |||
Graphic Display |
128×64 Graphic LCD | |||
Safety |
Built-in Smart GFI circuit,GFI trip current 450?A max, HV shut down speed: <1mS | |||
Key Lock |
To prevent unauthorized alteration of the test Parameters | |||
Calibration |
Build-in software and external calibrated meters | |||
Verification |
Build-in software driven verification menu to test fault detection circuits | |||
Alarm Volume Setting |
Range: 0-9 ;0=OFF, 1 is softest volume, 9 is loudest volume. | |||
Environment |
0-40℃, 20-80%RH | |||
Dimension/Net Weight |
215mm (W)×89mm(H)×370mm(D)/ 10Kg | |||
STANDARD ACC*SORI* | ||||
Power Cord(10A) |
×1 | |||
Fuses |
×2 (Including a spare contained in the fuse holder) | |||
Interlock Disable Key(1505) |
×1 | |||
Hipot Test Lead ,1.5m(1101) |
×1 | |||
Hipot Return Lead ,1.8m(1102) |
×1 | |||
*product specifications are subject to change without notice. |
7130
华仪(Extech)